10 Warning: Should it be determined that the air detection calibration failure has been caused by the presence of oxygen depleting compounds in t
11 c) When the hose is completely dry inspect the hose for signs of wear as described above. 3. Air Detection Sensor Replacement Inspection Fr
12 VI. Specifications Sample Parameters: Vapor only, oil and liquid free, 300 psig (2 MPa) Maximum Detected Compounds: R12, R134a, Air and un
2 Table of Contents I. Cautions and Warnings 3 II. Functional Description 4 III. Hardware Description 5 A. Instrument 5 1. Cas
3 I. Cautions and Warnings A. Cautions Never use any sample hose other than those approved for use with the ID900. The use of other hose type
4 II. Functional Description Contamination of refrigerant in air conditioning systems or storage cylinders can lead to component corrosion, comp
5 III. Hardware Description A. Instrument: The instrument houses the basic components required for operation of the ID900 as de
6 B. Sample Hoses 1. R12 Sample Hose: is a 6-foot hose consisting of a nylon inner tube surrounded by a polyurethane outer tube. The hose contain
7 IV. Operation Procedure A. Pre-Operative Inspection 1. Inspect the outside diameter of the white element contained within the sample filter f
8 5. When prompted by the instrument, connect the free end of the sample hose to the vehicle air conditioning system or refrigerant storage cyli
9 V. Maintenance A. Error Codes and Corrective Actions The instrument is supplied with self-diagnostic software to aid the user in the determina